Feature

New Cancer Surgical Tech Gets Positive Vote, But Some Cite Safety Concerns


 

Is the Technology Safe?

Pegulicianine is an imaging agent that contains a fluorescent dye. The agent is given to patients as a 3-minute intravenous infusion 2 to 6 hours before surgery.

After removal of the main tumor specimen, the surgeon inserts a handheld probe into the breast cavity and in combination with the detection software, searches for residual cancer that may have been left behind, Dr. Ferrer explained during the MIDAC meeting.

If the software identifies areas suspicious for residual cancer, those areas display in red on an overhead screen. The surgeon then takes a targeted shave to resect the suspicious tissue. Once the tissue has been removed, the surgeon can rescan the cavity with the probe to ensure a more complete resection has been performed. Use of the LUM system typically takes surgeons less than 7 minutes to use, Dr. Ferrer said.

In the study, a total of 406 patients received the intravenous pegulicianine, but 14 patients were withdrawn before randomization. After a standard lumpectomy procedure, 357 patients were assigned to the pFGS group and 35 patients to the control group.

Of the 406 patients, pegulicianine administration was stopped for adverse events in 6 patients (1.5%). Two patients had grade 3 serious adverse events related to pegulicianine; one had hypersensitivity, and one had an anaphylactic reaction. The other four pegulicianine-related adverse events included allergic reaction, milder hypersensitivity, nausea, and pegulicianine extravasation.

Dr. Schnabel said these reactions are worrisome. While any effort to reduce the need for patients to have more than one surgery to complete a breast conserving approach would be a “real advance,” Dr. Schnabel said she would not feel comfortable using pFGS in her own practice if approved by the FDA as is.

“This is clearly a major issue in terms of incorporating this technique into practice,” she said. “I could go on, but in light of the above, I’m surprised that [the committee] is supportive. I would hope for some refinement of the technique to reduce the risks to patients and improve the results before I’d consider utilizing this approach.”

During the MIDAC meeting, Dr. Ferrer said the company takes the safety events seriously and has developed mitigation strategies to further reduce the risk of patient hypersensitivity. These strategies include: clear labeling that informs users of anaphylaxis risk, incorporating a new section into the device training program to address warnings and precautions, an enhanced pharmacovigilance program to closely track and report hypersensitivity events, and a postmarket study to access the incidence rate and risk of such events in a broader population.

Several MIDAC members raised questions about the adverse reactions observed and about the safety of the technology.

David B. Hackney, MD, a neuroradiologist at Beth Israel Deaconess Medical Center in Boston, questioned the recommendation that patients only be monitored for 15 minutes after the injection.

“Since you don’t have enough data to know how long after injection reactions could occur, why not keep them under monitoring until after the surgery is over?” he said.

Barbara Smith, MD, PhD, lead investigator of the study, explained that per the protocol, there would be frequent monitoring, with a nurse at bedside, and patients would be monitored after injection, on their way to the procedure, and afterward.

She suggested, during the meeting, that more intense monitoring early in the process would be beneficial as that is when investigators observed side effects believed to be attributed to LUMISIGHT.

MIDAC member Kimberly E. Applegate, MD, a retired radiology professor, asked about the learning curve for surgeons and how long it generally takes for physicians to become familiar with the system.

Coinvestigator Kelly Hunt, MD, explained that all surgeons who participated in the trial completed a training program.

“Certainly, there’s a learning curve anytime we introduce new technology in the operating room,” she said. “Surgeons said it usually takes about three procedures before they’re comfortable with the system, including the camera and the software.”

During a presentation period by FDA officials, Anil Rajpal, MD, MPH, FDA, Deputy Division Director for Safety, said it’s important that prescribing information for LUMISIGHT communicate the risk of anaphylaxis and other hypersensitivity reactions, the need to monitor patients, and the need for the appropriate available personnel, medications, and equipment.

“This would be done by warnings and precautions and a boxed warning,” he said. “Note, that [such warnings] would only communicate the risks, it would not further characterize the risk.”

Recommended Reading

Magnetic Seeds or Guidewires for Breast Cancer Localization?
Federal Practitioner
Radiation Oncologists Fight for Payment Reform Amid Cuts
Federal Practitioner
Dana-Farber Moves to Retract, Correct Dozens of Cancer Papers Amid Allegations
Federal Practitioner
Chemo-Free Maintenance Strategies May Boost Survival in TNBC
Federal Practitioner
More Young Women Being Diagnosed With Breast Cancer Than Ever Before
Federal Practitioner
Commentary: Benign Breast Disease, PD-L1+ TNBC, and Exercise in BC, February 2024
Federal Practitioner
Hyperbaric Oxygen: Effective Against Cancer Radiation Harm?
Federal Practitioner
Unleashing Our Immune Response to Quash Cancer
Federal Practitioner
Is Mammography Ready for AI? Opinions Mixed on Usage, Cost Methods
Federal Practitioner
Does Exercise Reduce Cancer Risk? It’s Just Not That Simple
Federal Practitioner