FDA Class I recall: Batteries for CARESCAPE 2860 Ventilator


 

The Food and Drug Administration has issued a Class I recall for the CARESCAPE R860/Engstrom Carestation/Engstrom PRO Ventilator batteries. A total of 1,533 complaints allege that the batteries are draining much faster than expected, prompting manufacturer GE Healthcare to initiate the recall. There have been no injuries, and no deaths associated with the use of this device, according to an FDA corrected announcement.

Health care personnel and those patients who receive breathing support with these ventilators should be cautious about using CARESCAPE battery products moving forward, the agency said.

This type of ventilator is primarily powered by plugging into a wall outlet, but it has the capability to operate on backup batteries. These batteries are not solely for emergency situations such as power outages, but are also for routine situations such as transporting a patient within the hospital. GE Healthcare supplies these backup batteries with the ventilator, and sells replacements when they run out.

However, if the ventilator loses power because of battery malfunction, the patient may lose access to oxygen, leading to hypoxia, which can lead to brain injury and death. Therefore, if these batteries drain quicker than anticipated, it may put the patient at risk.

To prevent this danger, GE Healthcare recommends customers perform a battery performance test after they see this notice and every 3 months following. Consumers should take extra precaution and make sure their batteries are charged following a long period of inactivity. If the device is inactive for a while, the company says users should keep it plugged in to avoid draining the battery. Batteries should be replaced at a minimum of every 3 years.

When these devices are still plugged into the wall, they’re safe to use, according to the FDA. But when using the backup power source, clinicians should make sure to have alternative routes for breathing support on hand, such as with a bag-valve mask system.

There are 4,222 of these possibly defective batteries currently on the market. They were distributed from April 2, 2019, through April 18 of this year, when GE Healthcare stopped distributing these products and began the recall process. Any issues with these products should be reported to the FDA’s MedWatch database or by sending a medical device notification acknowledgment response to GE at the email address listed at the bottom of the recall announcement.

A version of this article first appeared on Medscape.com.

This article was updated 7/6/22.

Recommended Reading

Make America beautiful: Support mask mandates
MDedge Emergency Medicine
Should we always offer CPR?
MDedge Emergency Medicine
Doc accused of killing 14 patients in the ICU: Upcoming trial notes patient safety lapses
MDedge Emergency Medicine
SCAI refines cardiogenic shock classification system
MDedge Emergency Medicine
Analysis questions tocilizumab in ventilated COVID patients
MDedge Emergency Medicine
Hospitalists and PCPs crave greater communication
MDedge Emergency Medicine
ED staff speak out about workplace violence, ask for mitigation
MDedge Emergency Medicine
Surgeons, who see it up close, offer ways to stop gun violence
MDedge Emergency Medicine
What are the signs of post–acute infection syndromes?
MDedge Emergency Medicine
Childhood melatonin poisonings skyrocket in the past 10 years
MDedge Emergency Medicine