FDA/CDC

Deaths tied to reprocessed urologic endoscopes, FDA warns


 

The U.S. Food and Drug Administration is warning health care providers about the risk for potentially life-threatening infections associated with reprocessed endoscopes used for viewing the urinary tract, including cystoscopes, cystouerthroscopes, and ureteroscopes.

The federal agency is investigating more than 450 medical device reports, including three reports of deaths, received between Jan. 1, 2017, and Feb. 20, 2021, that describe post-procedure infections and other possible contamination problems associated with the reprocessing or cleaning and sterilization of the devices.

Although it’s early in the investigation, on the basis of available data, the FDA believes the risk for infection is low.

“We are very concerned about the three reported deaths – outside of the United States – associated with these infections, and we’re acting fast to communicate with health care providers and the public about what we know and what is still an emerging issue,” Jeff Shuren, MD, JD, director of the FDA’s Center for Devices and Radiological Health, said in a statement released on April 1.

Manufacturer Olympus Corporation submitted three reports of deaths attributed to a bacterial infection. In two of those reports, the infection was linked to a forceps/irrigation plug, an accessory component used to control water flow and enable access to the working channel of the endoscope. Lab tests confirmed that the bacteria that caused the infection was present in the forceps/irrigation plug.

The FDA said the third victim’s death involved a cystoscope that did not pass a leak test. It is possible that the damaged device was a factor in the patient’s becoming infected.

It’s not known to what degree the reported infections or patient comorbidities played a part in the patient deaths. The FDA also hasn’t concluded that any specific manufacturer or brand of these devices is associated with higher risks than others.

The FDA released recommendations for processing and using these devices and emphasized the importance of following manufacturers’ labeling and reprocessing instructions to minimize the risk for infection.

In addition to following reprocessing instructions, the recommendations include not using a device that has failed a leak test, developing schedules for routine device inspection and maintenance, and discussing the potential benefits and risks associated with procedures involving reprocessed urologic endoscopes with patients.

The newly reported concerns with urologic endoscopes are similar to problems associated with reprocessed duodenoscopes. In 2018, the FDA warned about higher-than-expected contamination rates for reprocessed duodenoscopes. The FDA has taken action on infections related to the reprocessing of duodenoscopes. In 2015, it required postmarket safety studies and the updating of sampling and culturing protocols. In 2019, the FDA approved single-use duodenoscopes in an effort to curb infections.

A version of this article first appeared on Medscape.com.

Recommended Reading

Large study finds trans men on testosterone at risk for blood clots
MDedge Internal Medicine
Surveillance after testicular cancer: New approaches slash radiation exposure
MDedge Internal Medicine
No vascular benefit of testosterone over exercise in aging men
MDedge Internal Medicine
BMI, age, and sex affect COVID-19 vaccine antibody response
MDedge Internal Medicine
Testosterone decline after steroid abuse revealed with new biomarker
MDedge Internal Medicine
Self-management techniques help relieve lower urinary tract symptoms
MDedge Internal Medicine
Is the WHO’s HPV vaccination target within reach?
MDedge Internal Medicine
PCOS equivalent in men: No ovaries required
MDedge Internal Medicine
New NAS report seeks to modernize STI paradigm
MDedge Internal Medicine
Risk for erectile dysfunction sixfold higher in men with COVID-19
MDedge Internal Medicine