FDA tweaks Impella indications on basis of postapproval study


 

The U.S. Food and Drug Administration has updated the Abiomed Impella RP System’s approved indications in a way that “better reflects the characteristics of the patients who may benefit the most from treatment with the device,” the agency has announced.

FDA icon Wikimedia Commons/FitzColinGerald/Creative Commons License

The revised language reflects the final results of a postapproval study in which survival rates for patients who met the premarket-study entry criteria were comparable to rates seen in the premarket studies, the FDA observed.

The postapproval study “further confirms that the device is safe and effective when used for the currently approved indication.” The indication’s added words, however, tighten the description of eligible patients in a way that more precisely reflects the premarket-study population.

The update states that the Impella RP System is “indicated for providing temporary right ventricular support for up to 14 days in patients with a body surface area ≥ 1.5 m2, who develop acute right heart failure or decompensation for less than 48 hours following left ventricular assist device implantation, myocardial infarction, heart transplant, or open heart surgery, without the presence of profound shock, end organ failure, or acute neurologic injury.”

The FDA “believes that when the device is used for the currently approved indication in appropriately selected patients, the benefits of the Impella RP System continue to outweigh the risks.”

The reworded indication is the latest among several updates to the agency’s February 2019 letter to clinicians noting a signal of increased mortality associated with the Impella RP device in an interim analysis of the same postapproval study. Ultimately, no such signal has emerged among the subset of postapproval patients who would have been eligible for the premarket study.

A version of this article first appeared on Medscape.com.

Recommended Reading

Fourth-gen transcatheter mitral valve shows clinical, procedural improvements
MDedge Surgery
In cardiogenic shock, edge-to-edge mitral valve repair improves outcome
MDedge Surgery
TAVR now used in almost 50% of younger severe aortic stenosis patients
MDedge Surgery
PASCAL for MV repair noninferior to MitraClip in CLASP IID; FDA took notice
MDedge Surgery
Angiography in patients with prior CABG does better when planned with CT
MDedge Surgery
Amulet, Watchman 2.5 LAAO outcomes neck and neck at 3 years
MDedge Surgery
Similar transplant outcomes with hearts donated after circulatory death
MDedge Surgery
ACC/AHA issues updated guidance on aortic disease
MDedge Surgery
Puzzling, unique ECG from pig-to-human transplanted heart
MDedge Surgery
In CABG, radial artery works best for second key graft: RAPCO at 15 years
MDedge Surgery