Plastic IUD placement instruments prevent uterine perforations


 

Key clinical point : In a biomechanical ex vivo analysis, metal uterine sounds caused uterine perforation, but the manufacturer’s plastic intrauterine device placement rod did not.

Major finding: The lowest mean maximum force generated for IUD placement was 12.3 Newtons with the levonorgestrel intrauterine system placement instrument, followed by 14.1 Newtons with the copper T380A intrauterine device placement instrument 14.1 Newtons; the highest mean maximum force of 17.9 N occurred with the metal sound ( P < 0.01).

Study details: The data come from 16 premenopausal women with benign conditions who provided hysterectomy sections at a single center.

Disclosures: The study was funded indirectly through grants to the University of Utah from Bayer, Bioceptive, Sebela, Medicines 360, Merck, and Cooper Surgical. Lead author Dr. Duncan had no financial conflicts to disclose.

Source: Duncan J et al. BMC Womens Health. 2021 Apr 7. doi: 10.1186/s12905-021-01285-6.

Recommended Reading

Etonogestrel implants may be bent, fractured by trauma or during sports
MDedge ObGyn
Does last contraceptive method used impact the return of normal fertility?
MDedge ObGyn
Oral contraceptives may reduce ovarian and endometrial cancer risk 35 years after discontinuation
MDedge ObGyn
Levonorgestrel IUD effective as emergency contraception
MDedge ObGyn
COVID-19 pandemic hinders access to contraception
MDedge ObGyn
How physicians can provide better care to transgender patients
MDedge ObGyn
Optimize your treatment of endometriosis by using an FDA-approved hormonal medication
MDedge ObGyn
Patient-centered contraceptive care for medically complex patients
MDedge ObGyn
Oral contraceptive with new estrogen earns approval
MDedge ObGyn
Oral contraceptive use shows no impact on later heart failure risk
MDedge ObGyn